Hagberg falling number measurement: New machine versus conventional machine

22/06/2018 09:22

PRESENTATION
Pre-harvest sprouting or germination on the mother plant (due to rain before harvest) and late-maturity alpha-amylase (due to heat or cold shock during grain development) cause alpha-amylase activity increase. The consequences of this for wheat flour and baked products can be very important and may lead to significant problems like sticky dough, bread with low volume and excessively red crust. The Hagberg falling number method was developed in the early 1960’s to provide a rapid means of determining the alpha-amylase activity. This widely accepted method measures the time required for a sensor to plunge into a heated flour and water gel.
 
Alpha-amylase present in the gel will cause it to degrade and reduce the viscosity of the gel. Thus, the plunger will fall faster. The shorter the time required for the plunger to fall indicates higher levels of alpha-amylase. The conventional FN machine (Perten Instrument) uses breakable glass tubes, boiling water and a cooling tower. A new device, Amylab FN (CHOPIN Technologies) using a safer induction heating system and reusable test tube made of aluminum is now available. The objective of this study is to evaluate accuracy performances of this new machine.
 

MATERIAL AND METHODS

20 wheat samples and 18 flour samples covering a wide range of alpha-amylase activity (from 90 s to 450 s) were analyzed both with the new Amylab FN machine and with the Perten FN machine.

 
RESULTS
20 wheat samples and 18 flour samples covering a wide range of alpha-amylase activity (from 90 s to 450 s) were analyzed both with the new Amylab FN machine and with the Perten FN machine.
The observed correlation coefficients (r2) between the two machines are equal to 0.98 for wheat and 0.98 for flour. In the same way, the difference between the two machines is lower than the uncertainty of ISO 3093 method for more than 95% of the tested samples.
 
CONCLUSION
This study confirms the Amylab FN ability to provide equivalent results to conventional machine with safer and more user-friendly conditions. 

 

Sightline Process Controls Inc., CHOPIN Technologies and Unity Scientific® to showcase baking process and quality control solutions at IBIE19

20190909

Las Vegas, NV (September 8, 2019) – Sightline Process Controls Inc., CHOPIN Technologies and Unity Scientific® – all KPM Analytics companies – will present their advanced analyzer solutions for baking applications at IBIE 2019, booths #4648 and #1953 at the Las Vegas Convention Center. 

阅读新闻

Examples for the application of dough mixing and heating behaviour in cereal research

20190620

How to differentiate between cereals varieties, to determine technological quality of grains or to understand the impact of components on the dough technological behaviour ? These are key questions, especially in the development of non-wheat and gluten-free products.

阅读新闻

CHOPIN Technologies inaugurates the AETC’s technical seminar series

20190613

The laboratory as a data source for making decisions » organized by AETC on may 8, 2019 in Spain. Learn more about our expertise in the cereals market, our motivations, our projects, the advantages of the Alveolab and our other devices.

阅读新闻